Assessing Non-Ideal Instrumental Effects in High-Resolution FTIR Spectroscopy: Instrument Performance Characterization

Article scientifique 2025 Anglais

Résumé

Abstract. This study investigates the impact of non-ideal instrumental effects on the performance of high-resolution Fourier Transform Infrared (FTIR) spectrometers, with a focus on the Bruker FTS 120M. Key non-idealities, including retroreflector misalignments, baseline drift, and spectral channeling, were systematically analyzed using advanced diagnostic tools such as ALIGN60 and LINEFIT. The nominal configuration exhibited significant anomalies, notably modulation efficiency (ME) deviations of up to +10.9 %, phase error (PE) variability of 2.11 × 10−2 radians, and spectral channeling frequencies such as a persistent 2.9044 cm−1, along with emerging frequencies around 0.24 cm−1 attributed to retroreflector wear and CaF2 beamsplitter degradation. A pronounced anomaly at 40.672 cm−1, likely induced by environmental factors such as external vibrations or mechanical instability, was also identified. Implementation of a modified configuration successfully mitigated these issues, reducing PE variability to 0.042 × 10−2 radians, aligning ME within the NDACC-acceptable threshold of 1.1, and achieving substantial improvements in the instrument line shape (ILS), including sharper peaks, narrower full-width at half maximum (FWHM), and reduced side-lobe asymmetry. Analysis of HBr transmission spectra revealed improved fitting of the P(6) line, characterized by lower residuals and enhanced spectral quality. Simulated Haidinger fringes near zero path difference (ZPD) highlighted alignment degradation patterns, underscoring the necessity for precise optical adjustments. Temporal trends showed a 14 % increase in ILS peak height and significant RMSE reductions in the modified configuration. Overall, this study provides a robust framework for diagnosing and correcting instrumental artifacts, ensuring the accuracy, reproducibility, and long-term stability of FTIR measurements essential for atmospheric trace gas retrievals.

Citer ce document

Daba, G., Tsidu, G. (2025). Assessing Non-Ideal Instrumental Effects in High-Resolution FTIR Spectroscopy: Instrument Performance Characterization. https://doi.org/10.5194/egusphere-2025-2232

Accès au document

Texte intégral en lecture en ligne, réservé aux abonnés SPHAERO et aux membres de l'institution. Se connecter

Voir l'article sur le site de la revue

Statistiques

Consultations : 1

Téléchargements : 0